Alcolizer Druglizer LE5 Gebruikershandleiding - Pagina 18

Blader online of download pdf Gebruikershandleiding voor {categorie_naam} Alcolizer Druglizer LE5. Alcolizer Druglizer LE5 20 pagina's. Charging cradle
Ook voor Alcolizer Druglizer LE5: Gebruikershandleiding (16 pagina's), Gebruikershandleiding (19 pagina's), Montage-instructies (3 pagina's)

Alcolizer Druglizer LE5 Gebruikershandleiding
3.5.8

Enabling Passive Only

To enable Passive Only, perform the same steps as in section 3.5.5. Press the Test key
until Passive Only is displayed.
A Passive Only test will not automatically enable an Active (Mouthpiece) test the same
3.5.9

Taking a Passive Only Breath Sample

The procedures for performing a Passive Only test are the same as those for performing
a Passive Standard Test (Refer Section 3.5.6.).
When a Warning Alcohol Present reading is displayed after performing a Passive Only
test, it is strongly recommended that a Mouthpiece Breath Test be performed, after the
3.5.10

Pressure Activated Passive Breath Test

If any step in a breath test sequence is not completed or the instrument is left idle for
longer than 60 seconds, the instrument will terminate the breath test and display test
timed out. The instrument will return to the press to test display.
If a test time out does occur during a breath test, a new breath test must be performed.
Perform a Pressure Activated Passive test as follows:
• Attach the supplied red Passive Cup to the sample inlet port of the instrument (Refer
Figure 19).
• On the Press Passive Test screen, press the Test Key. The instrument will perform a
background self-test while Please wait....is displayed on the screen.
REF: Alcolizer LE5 Alcohol Tester User Manual V15 docx
Figure 18- Warning Alcohol Present/ Press for Active Test Display
way a Passive Standard test does.
recovery period has ended (Refer Paragraph 3.5.6).
Note
Note
Note
Page 18
© Alcolizer Technology