TROUBLESHOOTING

In the event of a calibration failure, first check these common causes of errors:
• Plastic tubing has been overused, or has condensation present
• The alcohol gas standard value chosen does not match the gas being used or the value has
been incorrectly entered in the Select Value menu
• The alcohol gas standard is expired
• A connection to or from the gas valve is not secured properly (check all connections)
• The value entered in Altitude menu is incorrect
ALTITUDE REFERENCE GUIDE
The altitude adjustment is in increments of 200 meters. Choose the closest city elevation and
round the altitude up or down to the closest value accordingly.
EXAMPLE
CITY
Amsterdam
Berlin
Brussels
Geneva
Madrid
Moscow
Ottawa
Paris
Rome
Washington
SERVICE GUIDE
ALTITUDE (M)
0
34
48
373
654
189
79
34
14
7
SETTING
0
0
0
400
600
200
0
0
0
0
5