Direct Healthcare Group Bure Ortho S Technisch handboek - Pagina 8

Blader online of download pdf Technisch handboek voor {categorie_naam} Direct Healthcare Group Bure Ortho S. Direct Healthcare Group Bure Ortho S 12 pagina's.

Direct Healthcare Group Bure Ortho S Technisch handboek

5. Other instructions T-Motion

Control box wiring
Remove the control box cover to make the connections as described below. Use a screwdriver or similar; refer
to the special labels on the control box.
1. Socket for the raise/lower actuator. NOTE: Make sure that plugs are connected to the correct sockets
otherwise the actuators can be damaged.
2. Remote control socket.
3. Battery terminal socket.
4. Plug socket. NOTE: The plug must ALWAYS be connected to the control box for IP protection to be valid.
Battery charging
Connect plug to 230V wall socket. Charging begins automatically and the battery LED flashes green.
Make sure the battery is fully charged before using for the first time. It takes around 25 h to fully charge the
battery.
We recommend that the battery plug be removed from the control box if the product will be out of use for
an extended period (more than one week) or if the battery is switched off by means of the ON/OFF switch.
This eliminates the risk of disabling the battery as the control box always uses a small quantity of electricity
even when the walker is not in use.
1
2
3
4
LEDs on control box and battery
Control box – green light when power is available from battery/mains socket.
Battery:
The LED flashes green during charging.
ON/OFF: To use the product, press ON.
The green lamp will begin to flash.
Use the OFF button (hold down for 5 s) to switch off the battery if the product will not be used for an
extended period.
The LED will shows a constant green light when the battery is fully charged.
When the battery is in normal state/in operation, the LED flashes green at REGULAR intervals, namely 0.5
s ON, 4 s OFF.
8
BURE OR THO /O R THO S/SPACE
G ATE RD .SE