3. Troubleshooting

Warning / Fault
Cause
The Control
Control Unit does
Unit may not be
not operate;
attached to a
power source or
no display lights
a fuse may need
illuminate
replacing.
Mains failure /
Warning LED C
Other
+ audible warning
See above,
plus:
Pressure too low
Warning LED B
+ audible warning
Pressure too low /
Warning LED B+C
Air pipe kinked
+ audible warning
Pressure
Warning LED A
too high
+ audible warning
Alternating
Warning LED A+B
Mode Failure (no
+ audible warning
alternation)
Initialising Failure 1. Press the audible warning mute button to silence the audible warning.
Warning LED A,B+C
+ audible warning
6
D Y N A - F O R M  M E R C U R Y A D V A N C E T R O L L E Y
Solution
1. Check the Control Unit is connected to mains power outlet with the correct voltage.
2. Check the Control Unit is switched on. Switch off and unplug the unit before restarting.
3. Check the mains plug fuse (5 AMP) then check both Control Unit fuses (1 AMP) –
fuses can be released using a screwdriver to push and turn.
Do not try to open the Control Unit. Opening the unit could cause personal injury
or equipment damage.
Ensure the replacement of fuses is carried out accordance with local legislation.
1. Reset the audible warning – turn off power and press the audible warning mute button.
2. Check the handle is intact, ensuring all four sealing connectors are firmly fitted to the control unit
and the air hoses. Check the CPR tag is attached and all sealing connectors are firmly secure.
3. Check all air hoses along the inside of the mattress – each should be firmly connected.
Check each air cell is securely attached to its connecting air pipe.
4. Check all cells, pipes and hoses for any air leakage.
5. Switch on power.
1. Reset the warning – turn off power and press the audible warning mute button.
2. Check the handle is intact, ensuring all four sealing connectors are firmly fitted to the control unit
and the air hoses. Check the CPR tag is attached and all sealing connectors are firmly secure.
3. Check all air hoses along the inside of the mattress – each should be firmly connected.
Check each air cell is securely attached to its connecting air pipe.
4. Check all cells, pipes and hoses for any air leakage.
5. Check that the air filter cover is correctly secured and the air filter is clean.
6. Switch on power.
1. Check all air hoses along the inside of the mattress – each should be firmly connected.
Check each air cell is securely attached to its connecting air pipe.
1. Reset the warning – turn off power and press the audible warning button.
2. Disconnect the air hoses to reduce pressure, reconnect when pressure
has decreased.
3. Check for twists in the air hoses between Mattress and Control Unit.
4. Switch on power.
1. Reset the warning – turn off Power and press the audible warning mute button.
2. Disconnect the air hoses to reduce pressure – reconnect when pressure has decreased.
2. Check the power cable is firmly plugged into the mains outlet and the Control Unit;
and check the mains power is switched on.
3. Check the Control Unit fuse (1 AMP) – fuses can be released using a screwdriver
to push and turn.