Alcovisor Jupiter Gebruiksaanwijzing - Pagina 7

Blader online of download pdf Gebruiksaanwijzing voor {categorie_naam} Alcovisor Jupiter. Alcovisor Jupiter 12 pagina's.
Ook voor Alcovisor Jupiter: Handmatig (8 pagina's), Gebruiksaanwijzing (12 pagina's)

Alcovisor Jupiter Gebruiksaanwijzing

OPERATING INSTRUCTIONS

Pre-test Conditions
1. Operating temperature range is -10°C to 50°C. Use within this temperature
range will limit condensation from a subject breath to insure an accurate an
accurate reading from a breath sample.
2. It is recommended that the operator to observe a subject for about 20 minutes
before administering the breath test to insure that nothing is consumed or
expelled from the mouth of the subject.
eliminate reading from mouth alcohol.

Standard Test with Input

1. Attach a disposable mouthpiece onto the sample port.
2. Input information by a stylus pen.
NOTE: Maximum 16 characters for each input.
3. When displays "Please Blow", take a deep breath and blow until BEEP tone
stops.
4. Test result displays on the display.
5. Select
to print the test result.
6. Select
7. If testee did not blow within 30 seconds, tester can select "Refuse" to confirm
the testee refused to do the test; or select "Test Again" to allow testee to do
the test again.
8. If testee failed to give a valid breath sample, tester can select "Discontinue" to
confirm; or select "Test Again" to allow testee to do the test again.
9. A passive test can be performed by select "Passive" during the testee is
blowing.
10. Push the button underneath the mouthpiece to remove the mouthpiece.
to start another test.
Following this procedure will
P. 5