9. Technical Specification
Declaration – electromagnetic immunity
Guidance and manufacturer's declaration – electromagnetic immunity
The MAT1610001 is intended for use in the electromagnetic environment specified below.
The customer or the user of the system should ensure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge Immunity Test
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Power frequency (50Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
S E R V I C E M A N U A L
IEC 60601
test level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/output line(s)
± 1 kV line(s) to line(s)
Voltage Dips
Period
%U
(Cycles)
T
30
25
60
5
>95
0.5
Seconds
Voltage
Interruption
% U
T
>95
5
3 A/m
3 A/m
D I R E C T H E A L T H C A R E G R O U P . C O M
Compliance level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/output line(s)
±1 kV differential mode
Voltage Dips
Period
%U
(Cycles)
T
30
25
60
5
>95
0.5
Voltage
Seconds
Interruption
% U
T
>95
5
Electromagnetic
environment – guidance
Floors should be wood,
concrete or ceramic tile.
If floor are covered with
synthetic material, the relative
humidity should be at least
30%.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment.
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the Span system requires
continued operation during
power mains interruptions,
it is recommended that the
system be powered from an
uninterruptible power supply
or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.