Direct Healthcare Services Mercury Advance SMARTcare Servicehandleiding - Pagina 5

Blader online of download pdf Servicehandleiding voor {categorie_naam} Direct Healthcare Services Mercury Advance SMARTcare. Direct Healthcare Services Mercury Advance SMARTcare 20 pagina's. Pressure relieving mattress

Direct Healthcare Services Mercury Advance SMARTcare Servicehandleiding
LED Mode Settings
This symbol when illuminated (the green indicator light) is used to indicate that the equipment is
on or ready for use.
When a patient requires more pressure in the cells, as they may be uncomfortable or feel as
though the support surface is too soft or unstable, then please select a "High" setting (pressure
26mmHg). This must only be used by a trained clinician as often too high pressures can further
agitate certain patient conditions.
When a patient requires less pressure in the cells, as they may be uncomfortable or indeed hyper
sensitive to cell movement or if the patient is still reddening further, then please select a "Low"
setting (pressure 18mmHg). This must only be used by a trained clinician.
This function is used to silence the Audible Warning. The LED will remain lit if the Audible
Warning has been silenced previously, however a fault is still detected. Refer to the power switch
in order to re-set fully. If the Audible Warning continues to sound repeatedly, along with an
illuminated light, then an engineer must be called.
This symbol indicates an "Audible Warning Failure".
Please see troubleshooting guide below for how to re-set.
This symbol when illuminated indicates a Service is required.
DHG recommends a service every 8760 hours of operation (one continuous year running).
Note: Please ensure (when available) that all securing straps on the base of the mattress are secured onto
the MOVING PARTS of the bed frame.
For shut down procedure, see 4.2 Control Unit section.
S E R V I C E M A N U A L
D I R E C T H E A L T H C A R E G R O U P . C O M
Power On / Off
Dynamic /Firmer Setting
Low / Comfort Pressure Setting
Silence Audible Warning
Audible Warning Failure
Service Indicator
5