Warnings
Warnings
Read and follow all directions and warnings before using this device.
•
WARNINGS
1. F ailure to comply with all directions and warnings may result in injury or death;
use only as directed.
2. This device is designed for use on static mattresses only. DO NOT use this device
on alternating pressure or lateral rotation mattresses.
3. Due to the alternating pressure feature of the overlay, some devices and
products may not be appropriate for use with this device. Do not use pressure
pad alarm or alert systems in conjunction with the overlay.
4. This device is not suitable for all individuals. Other devices may be required.
5. I nspect the device for damage before each use and do not use if device appears
to be damaged or not functioning properly.
6. Never alter the device in any way.
7. T his device is designed for indoor use only within close proximity to
skilled caregivers.
8. T his device is not designed to operate with patients weighing more than 250 lbs.
NOTE: Resident's body can not exceed the width of the overlay.
•
WARNING – This device is designed to assist in the prevention and treatment
of pressure ulcers and may require other equipment. This may include, but is not
limited to:
1. Bed rails for repositioning and fall prevention.
2. Resident monitoring devices for elopement prevention.
3. Other devices as specified by the caregiver.
•
GROUNDING INSTRUCTIONS
Three-wire plug ground instructions
This device is for use on a nominal 120V circuit. This device should be grounded. In
the event of an electrical short circuit, grounding reduces the risk of electric shock by
providing an escape wire for the electric current. This device is equipped with a cord
consisting of a grounding wire with a grounding plug.
DANGER
Improper use of the grounding plug can result in a risk of electric shock.
If repair or replacement of the cord or plug is necessary, do not connect the
grounding wire to either flat blade terminal. The grounding wire is insulated with
either a green or green with yellow stripe outer surface.
Check with a qualified electrician or serviceman if the grounding instructions are not
completely understood, or if in doubt as to whether the device is properly grounded.
6