Nikon LABOPHOT-POL Gebruiksaanwijzing - Pagina 18
Blader online of download pdf Gebruiksaanwijzing voor {categorie_naam} Nikon LABOPHOT-POL. Nikon LABOPHOT-POL 24 pagina's. Polarizing microscope
Ook voor Nikon LABOPHOT-POL: Gebruiksaanwijzing (25 pagina's)
1;3. 'Monocular£yepiece Tube
66
Ap6I
\Pin
hole swing-in/out knob
\ ~Monocular
eyepiece
'--" tube "AP"
Bertrand lens
I~
in/out turret
Bertrand lens ~
focus turret
I
Fig. 25
1)
Bertrand lens
The Bertrand lens is brought in and out of
the optical path by turning
the Bertrand
lens tu rret.
The lens is in the optical path when the
indication on the turret is B.
The Bertrand lens can be focused by turn-
ing the focus turret
located under the
Bertrand lens turret.
2)
Pin-hole knob
The pin hole can be put in or out of the
optical
path by operating
the pin hole
knob located right-hand side of the eye-
piece sleeve.
By means of the pin hole, the conoscopic
image covering the area of
10J.1m¢
on the
specimen surface (when a 100
x
objective
is used) can be observed.
4. Epi-iliuminator
81M"
Used
for
episcopic
polarizing
microscopy,
mounted
between the Y -PO L stand and the
intermed iate tube "P".
1)
Nomenclature
and assembly
•
Referring to Fig, 26, assemble in the order
given.
CD
Remove the eyepiece tube and the inter-
mediate tube "P" from the Y -POL stand.
@
Mount
the epi-illuminator
"M"
on the
microscope arm, positioning
the illumina-
tor nearly parallel to the arm. Clamp the
screw.
®
After
releasing
sufficiently
the
clamp
screw on the lamp housing, to which the
lamp bulb
(6V-20W
Halogen lamp) and
socket is attached, insert the lamp housing
into the epi-illuminator
"M" and clamp the
screw.
@)
Connect the lamp cord to the transformer.
®
Place the epi-polarizer into the slot, facing
the white
index dot toward
the mirror.
[Note]
Insertion arid removal of the epi-
polarizer should be made with a
half-reflecting
mirror brought out
of the optical path.
®
Place the filters.
(j)
Mount the intermediate
tube "P" on the
illuminator
"M", positioning the tube with
its "Nikon"
plate faced to the operator,
regardless of the groove of the intermediate
tube "P" mount. Fasten the tube tentative-
Transformer
'4'
=-
•
~- Lamp vertical
centering ring
Lamp housing clamp screw
,-Socket sleeve clamp screw
--Lamp lateral centering screw
~l
'-----Lamp housing
"---Epi-illuminator
"M"
'-Half-reflecting
mirror
swing-in/out knob
"--"
~Slot
for epi-polarizer
"'-",-
L filter--
Intermediate tube
Il~
clamp screw -
I ~
-,
l(V
White index dot --
Epi-polarizer ------
NeB 10 filter
Intermed iate
tube "P"
Microscope arm
ND16 or GIF filter--
Fig. 26
18