Nikon N-STORM Bedieningshandleiding - Pagina 36
Blader online of download pdf Bedieningshandleiding voor {categorie_naam} Nikon N-STORM. Nikon N-STORM 48 pagina's. Super resolution microscope
Ook voor Nikon N-STORM: Gebruikershandleiding (11 pagina's)
9
Click
[Start STORM Analysis].
10
Select [Drift Correction], set [Periods] to [1], and then
click [Test].
Test analysis is performed for the currently displayed
period.
Upon completion of the analysis, the confirmation dialog is
displayed.
11
Click [OK].
Make sure that the bright points are identified correctly.
If the intended result is not achieved
If bright points of noise are detected or if bright points of
signals are not detected, reconfigure the parameters for
identifying bright points in [Identification Settings] and
then perform the test analysis again.
12
Click
[Start STORM Analysis] again.
13
Click [Start] to start analysis.
Analysis of bright points is performed for the entire
dataset. Upon completion of the analysis, the confirmation
dialog is displayed.
14
Click [OK].
Chapter 2
STORM Microscopy Operation
32
10
10
11
13
14