Nikon Nivo Series Handleiding communicatie-interface - Pagina 33

Blader online of download pdf Handleiding communicatie-interface voor {categorie_naam} Nikon Nivo Series. Nikon Nivo Series 34 pagina's. Hardware
Ook voor Nikon Nivo Series: Gebruiksaanwijzing (32 pagina's)

Nikon Nivo Series Handleiding communicatie-interface
Nikon Communication Interface Manual
6. Communication for SET type Data Recorder
There are three commands prepared for SET.
a) START command (11H): When an instrument receives the START command (11H) from
SET data recorder, it starts measuring and sends SD, VA, HA for every measurement until
receiving END command (12H).
b) END command (12H) : Cancel the START command.
c) DATA command (00H) : When an instrument receives the DATA command (00H), it
sends the latest data of SD, VA, HA regardless of current situation.
* In case of EDM OFF, DTM-500 sends the last measured data of SD, VA, HA after it receives
START or DATA command.
P33/34
Confidential