Select terminal input: select input terminal for signal input(option)
Exit: exit OSD menu

4.3 DICOM GSDF Calibration

The ARMSTEL Amedicom
DICOM GSDF compliant BARTEN curve for medical image display, with patented (pending)
TM
ARMSATT
technology embedded. It also allows user to push single "CAL" button to do the
automatic backlight adjustment and calibration. MDA2120-S is accurately DICOM Part 14
calibrated at the production. To adjust the display from shifting away from the DICOM
characteristic display function, ARMSTEL provides a dedicated tool for user to calibrate the
display lookup table whenever needed. The tool includes a calibration software package,
TM
Calispector
and a Simlux
TM
Calispector
software provides a user control function to calibrate the DICOM GSDF BARTEN
curve on site to prevent the shifting. The software allows user to calibrate at different levels of
accuracies, i.e., using 2048, 1024, 512, or 256 grayscales (grayscale luminance levels) to
interpolate the BARTEN curve. Using 2048 grayscales takes approximately 40 minutes to
restore the default setting, whereas 1024 takes around 20 minutes. 512 grayscales calibration
takes about 10 minutes while 256 grayscales takes less than 5 minute to complete the
calibration.
Calibration steps are as follows:
1) Install ARMSTEL Amedicom
following ARMSTEL website:
http://www.armstel.com/Amedicom®-dicom-calispector-software.php
The software supports Window 98/XP/Vista/Win7 OS;
2) Connect LCD display USB to the computer using the USB wire that comes with the LCD
display;
3) Connect ARMSTEL Simlux
TM
4) Fix the Simlux
meter on the LCD screen by itself with the sticky pat on the Simlux
®
MDA2120-A medical dual-mode display has the built-in calibrated
TM
meter compatible with the LCD display. ARMSTEL Amedicom
®
TM
Calispector
TM
meter to the computer;
-13-
MDA2120-A/MDA2120-S
software by downloading the software from the
User Manual
®
TM
meter.