Arometrix FRACTION FINDER Gebruikershandleiding - Pagina 4

Blader online of download pdf Gebruikershandleiding voor {categorie_naam} Arometrix FRACTION FINDER. Arometrix FRACTION FINDER 17 pagina's.

Arometrix FRACTION FINDER Gebruikershandleiding
● Light-Blocking Tape
● International Power Supply
● Adapter (optional)
● Warranty Card
Each FRACTION FINDER 2022 and newer comes with:
● Fraction Finder Sensor (Size 29 or 34)
● Arometrix Monitor
● Sensor Cable, USB
● Light-Blocking Tape
● International Power Supply
● Adapter (optional)

Section 4: Installation

The equipment should be in a clean, dry environment for the best results. Care should be taken to avoid
any spillage.
The Arometrix system was designed initially for short path distillation. Thus, the installation instructions in
this standard user manual focuses on how to operate the system for this specific application. If you are
using the system for another application, refer to our Application Notes on arometrix.com/resources.
1. Apply the light-blocking tape to the glassware apparatus, leaving an area for the sensor to be
installed. This is important as it will block ambient light from saturating your sensor's readings.
2. Install the sensor with the thicker part of the sensor down. The sensor should be installed on the
condenser just before the cow as shown.
(1) Having the glass be of high quality will help increase the signal to noise ratio – which will
yield better information.
(2) The sensor fits on most glass, but in some cases may require an adapter; visit our website.
Arometrix, Inc.
arometrix.com/resources
4