Diverse MF500M+ Gebruikershandleiding - Pagina 14
Blader online of download pdf Gebruikershandleiding voor {categorie_naam} Diverse MF500M+. Diverse MF500M+ 16 pagina's. Magmeter
Reading range
Occlusion
Technique
Accuracy
Zero
Probe orientation
Gravity
Display
Functions
Internal data sample rate
Repeatability
Probe size
Detection
Storage
Probe cable length
Calibration:
Power:
Instrument size:
Weight in case:
Environmental:
Humidity
Display update rate:
Warranty
MF500M Specification
0 to 2047 force units
0 to 100%
Uses net magnetic force to identify magnetite
and magnetic material
Reading range:
0 - 2047 force units. These are
arbitrary units, and each instrument needs to
be calibrated to the specific pipe geometry and
magnetite samples
Can be zeroed at any time - probe is corrected
for gravitational forces but zeroing
recommended in the measurement orientation
Measurement independent of probe orientation
provided it is zeroed first in that orientation
Probe is corrected for gravitational forces.
4 lines of 16 characters Shows value and user
interface messages
Raw force, occlusion for solid and broken
magnetite
500 samples per second
+/-5% assuming sample is degaussed
20mm x25mm, 150mm long
Detects the magnetic force between the
sensing magnet and the magnetite. The signal
is offset by an estimate of the gravitational
force.
Maximum of 20 records can be stored
1.5m typical
Calibration is indicative only. Unit is supplied
with a standard so that the instrument
functionality can be verified
4 standard AA cells , typical lifetime 12 months.
Continuous use 15 hours
210 x 100 x 50mm
1.3kg
10 - 40C operating, 0 - 80C storage
0-90% non condensing
0.3 seconds
12 months
Page 14