BALTECH OA-5000
Operation manual

4. Operating instructions

4.1 Calibration

Before calibration it is important to ensure the sensing cell is clean and dry, as
moisture and contamination will adversely affect the results. Testing should
generally be undertaken with the unit switched in the HI position. The
Green/Red scale is mainly for use when testing engine oils.
Step 1
Half fill the sensing cell with clean oil (ensure that the whole surface of the
sensor is well covered with oil) (Figure 4-1). Oil of the same brand as in the
system being evaluated, should be used for calibration.
Figure 4-1. Calibration step 1
Step 2
Run the rotary switch from the OFF position to the HI position (Figure 4-2),
the display will complete a full sweep and return to the zero position. Press
the CAL button for a minimum of 15 seconds (Figure 4-3). The display will
show with the CAL symbol flashing. If during testing the CAL graphic flashes
the unit will require re-calibration.
6
Operating instructions