Arthrex CONTINUOUS WAVE Snelle referentie - Pagina 7

Blader online of download pdf Snelle referentie voor {categorie_naam} Arthrex CONTINUOUS WAVE. Arthrex CONTINUOUS WAVE 8 pagina's. 4 set-up and operation

Arthrex CONTINUOUS WAVE Snelle referentie

TROUBLESHOOTING

PROBLEM
SOLUTION
1. Tubing sensor indicator is not seated
2. Ensure that the tubing sensor coupler
Tubing Out *
is seated completely
3. Change the tubing
* Return to Arthrex
Console fails
1. Ensure that no tubing is connected to the
Self Diagnostic
pump during power on sequence
Test*
* Return to Arthrex for repair
1. Check the AC main power cord
Console will
2. Try an alternate power outlet
not power up
3. Check the AC main fuses
* Return to Arthrex for repair
Lack of
1. Activate LAVAGE mode
distention liquid
2. Increase pressure
bloody/cloudy
3. Decrease outflow
1. Check for error messages
2. Open all tubing clamps and shut-off
valves
Does not pump
3. Ensure actual pressure is below the
when activated
target pressure
4. Check if the tubing is pinched, kinked,
or blocked
* Return to Arthrex for repair
1. Roller housing not secured – ensure
Door Open*
locking lever is properly secured
* Return to Arthrex for repair
1. Increase pressure
Inadequate
2. Activate LAVAGE mode
pressure
3. Reduce outflow
4. Use high-flow cannulas
For a full listing of warnings and instructions, please consult the Directions for Use and User Manual, packaged with the pump or
pump tubing. These materials can also be found on Arthrex.com.
for repair
®
PROBLEM
SOLUTION
1. Check for error messages
2. Check that all tubing clamps are open
3. Check the settings for flow and pressure
4. Check if the tubing is pinched, kinked
or blocked
No/Low Flow
5. Check that tubing seats correctly over
the rollers
6. Verify use of high-flow cannulas
7. If failure remains, return to Arthrex
for repair
1. If failure remains, increase or open outflow
Overpressure
2. Manipulate joint to stress-free position
1. Ensure adequate fluid supply
2. Decrease outflow
3. Check tubing for damage and if it is
pinched, kinked or blocked
Pressure Fault
4. Check tubing for proper connections
5. Replace tubing
6. If failure remains, return to Arthrex
for repair
Synergy Shaver Detection
1. Verify Synergy Resection Console is
at least 2.3
2. Ensure the Synergy cable is connected
No Shaver
3. Press stop and then run on CW4 while
Detection
the Synergy Resection Console is on
4. Replace the shaver detection cable
5. If failure remains, return to Arthrex
for repair