AimSafety PM100 Snelstarthandleiding - Pagina 2
Blader online of download pdf Snelstarthandleiding voor {categorie_naam} AimSafety PM100. AimSafety PM100 2 pagina's. Personal single gas monitor
Activation
Activating the Monitor
To activate the monitor, press and hold the [Function key] for 3 seconds. While the
key is depressed, a 3-second count-up timer will be displayed. Once the counter
reaches three (3), release the [Function key].
The monitor will perform the following startup sequence.
The unit will display the gas type, firmware version, display/alarm test, followed by a
10-second stabilization countdown. Once the countdown is complete, the monitor
defaults to Measure- ment mode, displaying the current gas readings and the icon.
Sensor readings may drift during shipping. All newly purchased monitors should be
bump tested to a known concentration of gas before use.
User Interface
Display Modes
Measuring Mode is the default mode. Once the monitor is activated the meter will
continually display measured gas concentrations in real-time.
Basic Mode is an optional mode where the remaining sensor life will be
displayed if the detected gas levels are below the alarm threshold. The monitor will
display the active gas reading only after an alarm threshold has been exceeded.
Basic mode can only be enabled using the PM Link and software, or the Bump Test-
Calibration Station.
Menu Screens
From the Display mode screen, pressing the [Function key] will step to the next
screen. Note: If you do not press the [Function key] within 10 seconds, the display
reverts to the main screen.
Stealth Mode - Press the [Function key] to advance to Stealth Mode (if enabled).
Stealth is an optional setting that disables all audible, vibrating alarms and alarm
LEDs. When Stealth is enabled, an "StL" screen is added to the menu to indicate
that the audible, visual, and vibrating alarms are silenced.
User Interface Continued
Peak MIN - Press the [Function key] to advance to Peak MIN indicated by the MIN
icon on the display (O
only), with the numerical value displayed. The Peak MIN is the
2
lowest concentration of oxygen that the sensor has detected since the peaks were last
cleared.
Peak MAX - Press the [Function key] to advance to Peak MAX indicated by the MAX
icon on the display, with the peak max concentration displayed. The Peak MAX is the
highest concentration of gas that the sensor has detected since the peaks were last
cleared.
Clear Peaks - Press the [Function key] to advance to Clear Peaks indicated by "CLr"
on the display. To clear the peaks, press and hold the [Function key] for three seconds.
The unit will beep once, and the MIN/MAX icon will turn off.
Remaining Life - Press the [Function key] to advance to Remaining Life indicated by
one of three icons on the display. Remaining life is the amount of time left on the
monitor before End-of-Life. The remaining life is displayed in months, days, or hours
as indicated by the display icons.
Alarm 1 - Press the [Function key] to advance to Alarm Set Point 1 indicated by the 1
flag on the display. This is the first (low for O
) set point that activates the monitors
2
alarms. The unit will store this alarm data in event logging.
Alarm 2 - Press the [Function key] to advance to Alarm Set Point 2 indicated by the 2
flag on the display. This is the second (high for O
) set point that activates the monitors
2
alarms. The unit will store this alarm data in event logging.
Firmware - Press the [Function key] to advance to Firmware Version indicated by the
V icon on the display. This is the current firmware version that is loaded into the
monitor.
Calibration - Press the [Function key] to advance to Calibration indicated by the
calibration cylinder bottle icon on the display. The calibration gas concentration will be
displayed. See Calibration for more information on monitor calibration.
Bump Test
A bump test should be conducted before each day's use.
Ensure that you are in a clean environment before performing a Bump
Test. The monitor must be calibrated if it fails a Bump Test.
To perform a manual Bump Test: (without the Bump Test Reminder active)
Ensure that the sensor is reading zero (or 20.9% for Oxygen)
•
Attach the calibration (cal) cap on top of the sensor inlet
•
•
Connect the hose from the gas regulator of the calibration gas bottle to the cal cap.
Ensure the calibration gas and gas concentrations matches the sensor installed in the
instrument.
Access the Calibration menu (see section "Accessing Calibration Menu")
•
•
"CAL" and
the icon are displayed.
•
Press the [Function key] once and "CAL" and the icon will appear.
Press the [Function key] again "btS" will be displayed.
•
Press and hold the [Function key] for 5 seconds and "tSt" will be displayed
•
Turn on the gas regulator
•
•
Once gas is detected, "GAS" will be displayed
• After the test is passed, "SUC" and the
icon will appear on the display followed by
the alarm notification and "btS"
•
Once the test has passed, remove the calibration cap and turn off the calibration gas.
The monitor will default to the main screen after 20 seconds or you can manually exit
the calibration menu (see Exiting the Calibration Menu)
Allow a few minutes for the gas to dissipate
•
•
After the gas has dissipated from the sensor, clear the sensor Peak values
If the sensor fails the bump test, an "FA" message with the
Alarm and Alert indication chart
Calibration
To access the calibration menu:
• Press the [Function key] to navigate to the Calibration screen.
• While the Calibration screen is displayed, press and hold down the [Function key] for 5
seconds to access the Calibration mode.
Once the Calibration menu is accessed, "CAL" and
menu has four options:
• Fresh Air Calibration
• Span Calibration
• Bump Test
• ESC – Exit Calibration Mode
To exit the calibration menu:
• Press the [Function key] until ESC is displayed.
• Press and hold down the [Function] key for 5 seconds to access the Calibration mode.
• Press the [Function] key again to return to the Measurement screen.
Note:
If you don't press the [Function] key within 20 seconds, the display reverts to the main
screen.
Fresh
Air Calibration
Fresh Air Calibration must be performed in a clean environment that is free
from other gases (calibration is assumed to be performed in an environment
icon will briefly display
with an Oxygen concentration of 20.9% Vol.). Fresh Air Calibration should not
be performed in a confined space.
To Perform a Fresh Air calibration
• Access the Calibration menu
• "CAL" and the
• Press and hold the [Function] key to start the Fresh Air calibration.
• A ten second countdown will appear on the screen.
Note: press the [Function] key any time during the countdown to abort.
• Once the countdown is complete "CAL", the
indicating a successful Fresh Air Calibration.
Note: If you do not press the [Function] key within 20 seconds, the display reverts to
the main screen.
If Calibration fails, the
the sales representative or AimSafety Technical Support.
Span
Calibration
When performing a Span Calibration only use certified calibration gas at the
required concentration level. Do not use expired calibration gas.
To perform a Span calibration
• Access the Calibration menu
• "CAL" and the
• Press the [Function] key once and CAL and the icon will appear.
• Attach the calibration (cal) cap on top of the sensor inlet.
• Connect the hose from the gas regulator of the calibration gas bottle to the cal cap
• Press and hole the [Function] key for 5 seconds to start the Span calibration.
• When calibration starts, a 90-second countdown displays.
Note: The countdown is only 60 seconds for 0
• Turn on the calibration gas.
• Once completed, the
• The device will return to Measuring mode.
• Turn off the calibration gas and remove the calibration cap.
• Clear the peak values for the sensor.
All alarms are muted for approximately 10 minutes after a successful Span
calibration.
If the calibration fails, the
calibration is performed. If calibration continues to fail, contact the sales representatives or
AimSafety Technical support.
the icon will be displayed. Calibration
icon will be displayed.
and
icons will be displayed
icon will appear on the display. If this continues, please contact
icon will be displayed.
2.
icon and the current gas measurement readings are displayed.
and icons and CAL will cycle on the display until a successful