CUSTOM ULTRASONICS, INC. 
9. Close the Storage Cabinet door. The Storage Cabinet is equipped with a lock. Routinely locking the
Storage Cabinet may reduce the incidence of unintended endoscope exposure or theft.
10. The Storage Cabinet's Adjustable Programmable Timer Dial is located outside of the Storage Cabinet
on the right front wall. Adjust the programmable timer to operate the air pump for the desired amount of
time (Up to thirty (30) minutes every hour). The Green LED is illuminated
during air pump activation.
a. The timer is factory pre-set to operate for ten (10) minutes per hour
and will remain off for the balance of one hour.
b. If set otherwise, the air pump will activate for the set time on the timer
and shut off for the balance of one hour.
Note: The air pump will only activate if the On/Off switch is in the
On position.
Note: The air pump will activate every hour, 24 hours a day as long
as the On/Off switch is in the ON position.
Note: Inside the Storage Cabinet on its right side is an electrical box. Located on the rear of
this box is a black momentary switch. When the black momentary switch is engaged, the
air pump will function. This switch is used for testing the operation of the pump.
11. Endoscope removal.
a. Open the Storage Cabinet door.
b. Remove the Storage Cabinet connectors from the applicable endoscope connections.
c. Remove the endoscope from the Storage Cabinet.

Daily Observations

1. Once a day, press the black momentary switch located on the back of the electrical box containing the
On/Off switch and timer.
a. When activating this switch, check to ensure that the air pump and exhaust vent fan are turned
on and are operational.
2. Once a day, listen for the operation of the air pump to confirm that it is coming on by the air purge timer.
a. The pump should operate for 10 minutes or the pre-set time noted on the label on the front of the
cabinet.
b. The pump is factory set at 10 minutes, but can be changed according to hospital/facility
requirements.

Daily Cleaning

1. The stainless steel drip plate [located on the inside of the Storage Cabinet base] must be cleaned and
sanitized daily using a damp lint free cloth and a standard hospital grade, non-abrasive, surface
disinfectant/cleaner.
MAN-009
Endoscope Storage Cabinet
Revision 08232021
Model 20000
Operator's Manual
Image: Timer Dial
Page 10 of 11