To calibrate, adjust the knobs on the resistance
3.
decade and start testing by touching both parts
of the touch button simultaneously
Table to evaluate the results
Accepted maximum
Calibration
deviation downwards
parameters
from expected value
Footwear system test
(lower limit)
Footwear system test
(upper limit)
Wrist strap/groundable
ESD smock system test
(lower limit)
Wrist strap/groundable
ESD smock system test
(upper limit)
The images in this document are for information only. We reserve the right to change the design, accessories,
technical specifications and various details of the product without notice. In case of improper use other than
as described in this documentation or in case of modification of the product which D és Tsa. Bt. has not given
its written consent, our company may reject any warranty claims.
Lower limit
value
- 20%
80 KΩ
- 10%
31,5 MΩ
- 20%
600 KΩ
- 10%
31,5 MΩ
15
Upper limit
Expected value
value
100 KΩ
120 KΩ
35 MΩ
38,5 MΩ
750 KΩ
900 KΩ
35 MΩ
38,5 MΩ
Accepted maximum
deviation upwards
from expected value
+ 20%
+ 10%
+ 20%
+ 10%