Aseptico Sealant Express ADU-08SXE Bedieningsinstructies - Pagina 3

Blader online of download pdf Bedieningsinstructies voor {categorie_naam} Aseptico Sealant Express ADU-08SXE. Aseptico Sealant Express ADU-08SXE 4 pagina's.

Aseptico Sealant Express ADU-08SXE Bedieningsinstructies
ADJUSTMENTS:
Master Pressure-
supply pressure. It is located behind the vacuum holder, underneath the
control panel, inside the unit. This regulator is factory set and should not
require further adjustment. If further adjustment is needed, turn knob
clockwise to increase pressure and counter clockwise to decrease.
Water Pressure-
the water supply bottle. It is located to the left side of the water bottle, under-
neath the control panel, inside the unit.
knob clockwise to increase pressure and counter clockwise to decrease.
Syringe Flow-
syringe. It is located directly behind the syringe, underneath the control
panel, inside the unit. Using a small screwdriver blade, turn the right
adjusting screw clockwise to decrease air flow and counterclockwise to
increase.
Handpiece Drive Air-
It is located directly behind the syringe, underneath the control panel, inside
the unit. Using a small screwdriver blade, turn the left adjusting screw
clockwise to decrease drive air pressure, and counterclockwise to increase.
The pressure gauge on the front panel indicates handpiece running pressure.
Water Coolant Flow-
flow to the handpiece control. It is located directly behind the handpiece
holder, underneath the control panel, inside the unit. Turn the knob clock-
wise to decrease water coolant flow, and counterclockwise to increase.
SEALANT
MASTER
REGULATOR
MASTER
PRESSURE
ADJUSTMENT
The master pressure regulator controls the incoming
The water pressure regulator controls the pressure inside
The flow block controls the air flow to the three-way air/water
The flow block controls the drive air to the handpiece.
The water coolant flow knob controls the water coolant
ADU-08SXE CONTROL PANEL
WATER
REGULATOR
WATER
PRESSURE
ADJUSTMENT
If further adjustment is needed, turn
HANDPIECE PRESSURE
WATER
COOLANT
ON - OFF
FLOW
WATER
BOTTLE
MOUNT
H.P. WATER
COOLANT
FLOW ADJUST
ADU-08SXE
AIR ADJUST
H.P. DRIVE
SYRINGE
AIR
AIR
SYRINGE AIR
ADJUSTMENT
HANDPIECE DRIVE
AIR ADJUSTMENT