Flammable Environment
2.1.3.1 Some materials, for example cotton, wool, and gauze when saturated with oxygen may
be ignited by sparks produced in normal use of the equipment.
2.1.3.2 Do not use the Generator in the presence of flammable anesthetics or other flammable
gases, flammable liquids, or flammable objects.
2.1.3.3 Do not use this device in oxygen enriched atmospheres or in the presence of nitrous
oxide (N 2 0).
2.1.3.4 Use only non-flammable agents for cleaning.
2.1.3.5 Flammable agents used as solvents for removing adhesives should be allowed to
evaporate before powering on the Generator. Any flammable fluid pooled under the
patient or in the body depressions should be mopped up prior to using
the Generator. Attention should be called to the danger of ignition of endogenous gases.
NOTE: To ensure safe and reliable operation, the Radiolase
accessories should be checked for wear and tear before operation.
A maintenance contract with the manufacturer or one of its authorized service agents is
recommended after the warranty period.
3.0
SUPPLIER'S INFORMATION
Cynosure, Inc. dba Ellman
400 Karin Lane
Hicksville, New York 11801 U.S.A.
(516) 594-3333 • (800) 835-5355
Supplier will make the following available upon request; circuit diagrams, component parts list,
description, calibration instructions or other information necessary to assist technical personnel.
ATTENTION: Alcohol must never be used to clean and disinfect the front panel.
3.2
CLEANING, DISINFECTING AND STERILIZING THE UNIT AND ACCESSORIES
3.2.1
Use only nonflammable and nonexplosive agents. No moisture or detergents should
enter the unit.
3.2.2
Handpiece and Electrodes.
Always clean and disinfect the electrodes prior to sterilization.
Disinfect and sterilize handpieces and electrodes according to the directions supplied
with them.
All autoclavable products should be routinely inspected for any wear, breaks, or
deterioration of the insulation, and should be discarded and replaced if any of these
occurrences exist.
3.2.3
The footswitch is a waterproof unit; it can be cleaned with regular detergent.