ALARMS
If there is a pressure alarm, the red indicator will flash and an intermittent audio signal will sound. The audio signal can be cancelled
by pressing the
button.
1.
Check the feed pipe connection to the compressor and that the CPR valve is closed
2.
Check that the air lines are positioned and connected correctly and not split
3.
Check that there are no leaks in the cells (structure, connection)
4.
Replace defective parts if necessary (pipes, cells, etc.)
5.
If the problem persists, contact your reseller
When the pressure returns to normal, the compressor will revert to normal operation and the indicator will go out.
If there is an electrical fault alarm, the red indicator will flash and a continuous audio signal will sound. The audio signal can be
cancelled by pressing the
1.
Check that the compressor is still connected (to the pump and to the power point) and that the mains indicator is on.
2.
If the problem persists, we advise you strongly to disconnect the air lines, blank them off with the transport plug and move the
patient regularly.
When the power supply returns, you must restart the compressor by pressing the Start/Stop button while repeating the
INSTALLATION/USE instructions, §.I, starting from step 8, after you have reconnected the mattress feed pipe to the compressor.
COMMENTS:
-
On starting the compressor, there is an initialisation step. If there is a fault, an alarm is indicated by rapid flashing of all the
indicators or, depending on the type of problem by the servicing indicator flashing.
- If the compressor is disconnected, even temporarily, there is always an electrical fault alarm to warn against accidental handling.
WARNING:
After accidental deflation of the mattress due to a prolonged power cut, check at the time of reflation that the patient does not have a
limb wedged between the mattress and the bed sidebars.
CPR (Emergency Cardio-Pulmonary Resuscitation)
In case of cardiac arrest:
1.
Turn the end of the CPR valve to set the dial to OPEN,
2.
The mattress will deflate and the back of the thorax will be in less than 15 seconds on the firm base of the bedframe to allow
External Cardiac Massage.
button.
- 9 -