Monitor 500 – Quick Manual
During measurement pay attention to
Contralateral
CNAP
During measurement of interarterial BP
the CNAP finger cuff MUST NOT be
attached to the same arm.
- Selection of the correct size of NBP cuff and CNAP finger cuff
- Induction must not start before CNAP
- Optimal: NBP interval = 15 min, CNAP interval = 30 min
- Positioning of patients: Put no external pressure on the CNAP finger cuff
- On relocation check position of finger cuff and trigger manual change finger.
If required stop measurement, reposition cuff and restart measurement.
- On repositioning of the arm with the finger cuff: trigger NBP
- Artifacts at end of OP (repositioning ...)
- Stop CNAP measurement when patient wakes up (movement artifacts)