Coaguchek XS Plus System Pierwsze kroki

Przeglądaj online lub pobierz pdf Pierwsze kroki dla Sprzęt medyczny Coaguchek XS Plus System. Coaguchek XS Plus System 2 stron.

CoaguChek
System
setup
3. set the language
english is the default selection. It is
highlighted in black.
Press
to save your language setting.
The display changes to the Date and
Time Format screen.
preparIng FOr a test
1. gather the Following Items
• CoaguChek XS Plus Meter
• Container of test strips
• Test Strip Code Chip
• Accu-Chek Safe-T-Pro Plus
lancing device
• Alcohol wipe
• Cotton ball
testIng
1. remove a test strip
Take a test strip out of the container.
Close the container tightly.
You have 10 minutes to use a test strip
once you remove it from the container.
6. record result
Record your result as instructed by
your facility.
Note: If during testing the meter
displays an error message, refer to
the Error Messages section of the
Operator's Manual for an explanation
and steps on how to proceed.
XS Plus
®

getting started

4. set the Date and time Format
MM.DD.YYYY is the default date
selection.
12H is the default time selection
Press
to save your date and time
format settings.
The display changes to the Date screen.
2. Match Code
The code number on the test strip
container and the test strip code chip
must match.
Each box of test strips comes with a
matching code chip. Test strip code
chips are identified with an "S."
Every time you open a new box of test
strips, you will replace the test strip
code chip.
2. Insert test strip
Slide the test strip into the test strip
guide in the direction of the arrows
until it stops.
The meter powers on, and the main
menu appears.
If you insert a strip from a new lot
number for the first time, the meter
will display the number of the test strip
code chip required. Remove the old
code chip and insert the new one.
7. Clean up
Place the used strip and lancet in an
approved biohazard container.
Power the meter off.
Follow these steps to get started using the meter:
1. Watch the CoaguChek XS Plus System DVD. It will help you get comfortable with the
CoaguChek XS Plus Meter and the testing procedures.
2. Read this Getting Started guide. You will learn how to set up and perform your first
blood and optional liquid quality control tests on the CoaguChek XS Plus Meter.
3. With this Getting Started guide by the meter, follow the steps to perform your first test.
then, as necessary, refer to the Operator's Manual:
The CoaguChek XS Plus System Operator's Manual is a comprehensive guide to the
meter and test strips. It is designed to provide answers to your questions about the
meter' s operation and use.

InstallIng BatterIes

1. Open Compartment
Press the release tab on the battery
cover on the back of the meter.
Slide the battery cover off.
After you insert the batteries, the meter powers on and the screen changes to the
language menu.
You'll need to set several parameters before using the meter for the first time.
5. set the Date
Date
09:15 am
2012
Year:
1
Month:
1
Day:
04/19/2012
or
Press
to select the year, then
or
to select the month, then
or
to select the day.
Press
to save the current date.
The display changes to the time screen.
The date settings are important. Each time
you run a test, the meter compares its date
with the test strip's expiration date. An
incorrect date may cause an expired strip
error on in-date strips preventing you from
running a test.
If the test strips are expired, the meter
displays an error message and prevents
you from running a test.
3. Insert test strip Code Chip
With the code number facing up,
insert the code chip into the code chip
slot until it snaps into place.
3. select patient test
Press patient test on the main
menu screen.
Enter the Patient ID.
An hourglass symbol appears.
A test strip symbol appears with a
flashing blood drop. You have 180
seconds to apply the sample.
2. Insert Batteries
Insert 4 AA batteries according to
the diagram inside the battery cover.
Replace the cover.
6. set the time
or
Press
to select the hour, then
or
to select the minutes
then aM or pM.
Press
to save the current time.
The display returns to the Main Menu.
4. Wash Hands
Have the patient wash his or her
hands in warm, soapy water. Or, clean
the fingertip with an alcohol wipe.
Make sure the fingertip and your
gloved hands are thoroughly dry.
4. Collect the sample
Twist the protective cap off the lancet.
Massage the finger until you see
increased color in the fingertip.
Keeping the hand down, press the
tip of the lancet firmly against the
side of the fingertip. Press the blue
trigger button.
Gently squeeze from the base of the
finger to develop a drop of blood.
Screen
09:15 am
Contrast
Setup Menu
09:15 am
Setup Menu
09:15 am
Result Units
Screen
Screen
Language Selection
Screen
09:15 am
Options
Options
Contrast
Date / Time
ID Setup
ID Setup
Result Units
You may change or re-enter the date,
09/16/2009
Lockouts
Lockouts
time or format at any time. Always access
Language Selection
these options by:
09/16/2009
Pressing setup on the main menu, then
09/16/2009
Date / Time
screen on the setup menu, then
Date/time on the screen menu
09/16/2009
...which brings you to the Set Date,
Set Time, and Format options.
Before continuing, review these
tips for getting a good blood drop.
Increasing the blood flow in the finger
will help you get a good drop of blood,
so keep in mind these tips:
• Warm the hand. Have the patient
hold it under his or her arm, use
a hand warmer, and/or wash with
warm water.
• Have the patient let that arm hang by
his or her side.
• Massage the finger from its base.
Use these techniques until the
fingertip has increased color.
5. apply the sample
Within 15 seconds of sticking the
fingertip, apply the blood to the target
area on the test strip.
Find the clear target area on the
test strip. You can dose from the
side or top.
Hold the blood drop to the test strip
until you hear a beep. The flashing
blood drop symbol will disappear and an
hourglass symbol symbol appears and
the test begins.
Do not add more blood to the test
strip. Do not touch the test strip
while the test is in progress.
The meter performs an automatic quality
control check on the strip before it
displays the test result.
appears in the display briefly
QC
following a successful QC check.
The result appears in less than a minute.