Section 4. Parameter Setup

The DS4100 offers the ability to change the shutoff timer settings. Parameter settings
shown in bold are the default settings.

Entering and Exiting Setup Mode

To enter setup mode from the front panel, press and hold REWEIGH and then
press and hold ZERO. Release both buttons when the display reads CAp 25.
To exit the setup mode, press KG until the display returns to the weigh mode.

Setup Mode Navigation

1. Press KG to change the current parameter.
2. Press LB to change the parameter setting.

Calibration Weight Selection

Cal
25 (10 kg)
45 (20 kg)

Calibration Mode Selections

CAL
0
CAL
25
45
25 LB (10 kg)
45 LB (20 kg)
Zero Calibration Point
Press ZERO to calibrate zero point
NOTE: This parameter is only visible when
in the calibration mode.
Span Calibration Point
25 LB
45 LB
NOTE: This parameter is only visible after a
zero calibration is performed.