DAESUNG MAREF DOCTOR LIFE III DL2002D Podręcznik - Strona 14
Przeglądaj online lub pobierz pdf Podręcznik dla Sprzęt medyczny DAESUNG MAREF DOCTOR LIFE III DL2002D. DAESUNG MAREF DOCTOR LIFE III DL2002D 19 stron. Compressible limb therapy system
How to connect 4 hose jacks to a cuff
7)
Put 4 hose jacks to 4 caps of the same color on cuff!
The effect of therapy will drop if the order of putting is not right.
Please check definitely the order of pressurization is applied from foot→calf→knee→thigh in turn.
8) How to use one leg cuff
9) How to use two leg cuffs
- As side picture, wear a leg cuff to user's leg and put
1-line hose plug perfectly to air socket of a main body.
If not so, air can leak from air socket.
- You can just use 1-line hose in using only one cuff.
- As side picture, wear two leg cuffs to user's two legs and
put 2-line hose plug perfectly to air socket of a main body.
If not so, air can leak from air socket.
- You can just use 2-line hose in using two cuffs.
14