Baumer 09 Series Instrukcja obsługi - Strona 6

Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Akcesoria Baumer 09 Series. Baumer 09 Series 11 stron. Ultrasonic sensors rs-232

Load factory
D
settings
Set measuring
A
mode
Set output
F
format of
permanent
periodical data
output
Set sensor
B
sensitivity
(Sensors with
beam
columnator)
Set number of
C
averagings
Temperature
G
compensation
on, off
Teach-in of Sdc X
Bedienungsanleitung_Series09-US_RS232_V1_e.doc
26.03.2013 14:43:00/tof
Configuration
{0D}
{0Dßß}
{0AX}
{0AXßß}
{0FX}
{0FXßß}
{0BX}
{0BXßß}
{0CX}
{0CXßß}
{0Gx}
{0Gxßß}
{0X}
{0XXßß}
The Sensor is configured with its
factory settings.
See chapter „Factory
configurations"
X
A: absolute measurement in 0.1mm
steps
B: relative measurement within the
Teach-range 0...4096
See chapter „Commands explained
in detail"
X
A: ASCII
B: Binary
See chapter „Commands explained
in detail"
X
A: Highest sensitivity
Measuring into test tubes
Measuring range: 3...150 mm
B: Second highest sensitivity
Measuring into plates size 96 with
big hole diameter
Measuring range: 3...110 mm
C: Second lowest sensitivity
Measuring into plates size 96 with
small hole diameter
Measuring range: 3...70 mm
D: Lowest sensitivity
Measuring into plates size 384
Measuring range 3...30 mm
See chapter „Commands explained
in detail"
X
A: not averaged
B: 2
C: 4
D: 8
E: 16
F: 32
G: 64
See chapter „Commands explained
in detail"
x
0: Temperature compensation off
1: Temperature compensation on
See chapter „Commands explained
in detail"
X
Reads and stores close limit Sdc
A: Teach-in successful
B: No object within measuring
range → basic setting for the
sensitivity the currently set
sensitivity level
6/11