3. 0   Important Safety Precautions 
The CDN must be securely fastened to the earth ground before making any connections to 
external power. Proper grounding of the CDN is required in order to prevent high leakage 
currents from presenting a potential safety hazard for test personnel.  To help establish 
proper grounding, the bottom surface of the CDN is unpainted. In addition, the bottom plate 
has mounting holes for securely bolting the CDN to a ground plane. If proper grounding has 
been achieved, the resistance measured between the CDN bottom plate and the ground 
plane should be less than 10 milliohms.  
 
Do not connect the power to the CDNs until all the test equipment and EUT are in place and 
electrically connected.  Avoid having any exposed live wires or connections.  You can use the 
following procedure.  
 
1. Connect EUT cable to CDN EUT port. 
2. Connect the Test Generator (signal generator, power amplifier, attenuators, etc) to 
the RF port of the CDN. The test generator output must remain disabled until all 
connections are made. 
3. Connect accessory equipment (AE) port. 
3.1 
Maximum RF input  
Do not exceed RF input level to the CDN. Excessive RF input will damage the CDN input and 
will not be covered under warranty. To avoid accidental damage to the power amplifier or 
CDN, disable the RF output of power amplifier before making any connections. Also make 
sure the signal generator output is set to its minimum level when the RF is turned on. 