3. Measuring principle

This module monitors by Constant incident light
Irradiated
on liquid surface,The receiving chip will dynamically measure
the chroma of water 。 If the water is turbid, it may affect the
measurement accuracy , It is recommended to filter and clean
before entering water。

4. Structural features

Outline structure and installation instructions of
Meter,
Online Color
As shown in the figure below。
Careful:
1. Outflow valve flow control to minimum, It must be
less than the inflow flow , Let the measuring tube
have a certain air pressure,Otherwise, there will be
bubbles,Influence measurement value
Panels should be installed vertically
2.