Notes on Displacement Power Factor & True Power Factor
●Introduction: Power is the rate of change of energy with respect to time (in terms of
voltage V and current A). Instantaneous (real) power w = vi where v is the instantaneous
voltage and i the instantaneous current. The average (real) power is the mean of vi and
is given by:
W = ω/2π∫vi dt , over the interval from 0 to 2π/ω
●Displacement Power Factor (more traditional): Assuming V and A are pure
sinusoidal waveforms without harmonics (as in most traditional cases), that is, v = V sin
ωt and i = I sin (ωt -θ), the expression can be simplified to:
W = 1/2 x V x I x Cosθ where V and I are the peak values, θ is the displacement
power factor angle, and Cosθ is the displacement power factor. Using RMS values, it is
written as:
W = V
x A
rms
Practically, in such cases without harmonics, θ is also called the phase-shift angle of
the current A to the voltage V. An inductive circuit is said to have a lagging power factor
since current A lags voltage V (phase-shift angle θ and thus Sinθ are both "+"), and a
capacitive circuit is said to have a leading power factor since current A leads voltage V
(phase-shift angle θ and thus Sinθ are both "-").
●True Power Factor (encountering harmonics): When encountering distorted
waveforms with the presence of harmonics, however, the simplified power expression
should not be used since substituting the above mentioned pure sinusoidal V and A
functions cannot fulfill the actual conditions. Cosine of phase-shift angle (Cos
displacement power factor, is no longer the only component constituting the overall
power factor. Harmonics do increase apparent power and thus decrease the overall
power factor. That is, the True Power Factor is actually affected by both phase-shift
angle and harmonics, and is given by the expression:
True Power Factor (PF) = Real Power (V
A
)
trms
In order to improve overall system power factor, nowadays power-system engineers
need to address both phase-shift and harmonics problems. Practically, harmonics
should be dealt with (e.g. filtering out) before phase-shift to be corrected (e.g. installing
capacitors in parallel with inductive loads).
Power function
Set the slide-switch function-selector to the Power position.
●Default at last selected function. Press SELECT button momentarily selects between
W (real power), VAR (reactive power) & VA (apparent power) measurement functions.
x Cosθ
rms
x A
x Cosθ)/Apparent Power (V
rms
rms
7
), or the
θ
x
trms