Function Key Menu

No changes to controller default parameters.

Global Menu

Parameter
Display Units
AC Line Frequency
Zone
Channel

Communication Menu

Parameter
Standard Bus Address

Troubleshooting

Symptom
Chamber does not control
temperature and humidity
Humidity does not increases
fast enough, or does not
increase above 75%RH
Humidity does not decrease
fast enough
H300 Temperature Humidity Chamber
Display




Display

Causes
The CONT switch is in the OFF position.
1. The water in the water reservoir is too low
2. The water in the water reservoir is too cold. Allow about 30 minutes
for the water to warm up to normal operating temperature.
3. The chamber door is not properly sealed
4. The water in the reservoir is not distilled.
1. Desiccant is used up (red). Make sure some of the desiccant in the
desiccant reservoir is blue.
2. Condensation in the chamber.
3. The chamber door is not properly sealed.
Temperature
Humidity






Temperature
Humidity