Alarms and Troubleshooting
Flashing red LED - High priority alarm
Issue
Solution
Charge the battery by connecting the AC adapter to
the pump and power source.
Note: After plugging into the power supply, the
user has to restart the feeding by pressing the
multifunction button [ ].
Release upstream occlusion.
X
Make sure the tubing is positioned correctly.
If there is an occlusion between the pump and the
patient, release the occlusion and ensure that the
branch of the cassette is empty.
If alarm is not resolved, check that the cassette is
positioned correctly between the two black pins and
that there is no presence of feeding solution between
the two black pins.
If the alarm is still not resolved, replace the tubing.