1. Take the stand to the working place and arrest the two rollers with brakes.
Caution: Do not touch damaged electrical parts. Danger of electrical shock!
If you notice a damaged mains cable or plug do not plug in. Call your service number
and report the damage.
2. Check if mains plug or mains cable are damaged.
3. Please check that an earthed socket is available within the working area of the lamp. The
socket to be used for the lamp must be installed according to the requirements stipulated by
the IEC or VDE 0107.
Remark:
Do not overrun any objects on the floor or the mains cable. The stand might tip over.
Note: Before doing any cleaning work turn off the lamp, draw plug from mains and make
sure that the lamp cannot be switched on again.
Please ensure, that no disinfection liquid flows into the joints or lamp head.
The surface of the stand can be easily kept clean by simply wiping with a wet cloth. You can
use conventional cleaning agents.
Caution:
In order to prevent any damage at plastic parts do not use scouring agents or alkaline, acidic
and alcoholic cleaning agents.
For cleaning proceed as follows:
•
Draw plug from mains.
•
Wipe the surface of the stand with a wet cloth. As a cleaning agent you can use water or a
soap solution (washing-up liquid).
Note: Before doing any maintenance work, turn off the lamp, draw plug from mains and
make sure, that the lamp cannot be switched on again.
The following maintenance work / tests has / have to be done every six months:
•
check on defects in paint work;
•
check on fissures at plastic parts;
•
check on deformation of the suspension.
The following maintenance work / tests has / have to be done once a year:
•
check the function;
•
electrical safety test;
•
check the suspension;