Drawell DW-2002H Instrukcja serwisowa - Strona 8

Przeglądaj online lub pobierz pdf Instrukcja serwisowa dla Mikroskop Drawell DW-2002H. Drawell DW-2002H 11 stron. Epi-fluorescent microscope

1. The Principle:
The light source of the Epi-fluorescence device will eradiate light with specific wavelengths.
It will be transmitted on the sample and excite the sample to produce fluorescence. You may have a
high brightness and high contrast image to achieve fluorescence observation and analyze.
2. Function:
The range of the spectrum excited: 350nm ~ 580nm
The range of the fluorescence spectrum: 420nm ~ 650nm
3. Technical Parameters and Outfit:
Name
Epi-fluorescence
device
Microscope body
Viewing head
Eyepiece
Objective
Power supply box
Light source
Power cord
Immersion oil
Ultraviolet bar
Service manual

D) How to Use the Fluorescence Microscope:

1. Turn on the switch (17) of the power supply box (19). It will be in full working condition
after 10 minutes. The tension-guide line (16) will show the state of the mercury lamp.
2. Adjusting the center of the mercury lamp:
Push the excitation type selector (5) to the position "o", and push the flashboard B (6) in a
middle position. You will find a light shadow on the observation window (21). If the light shadow
is not in the middle of the window, rotate the screw (8) to move the shadow up/down, rotate the
screw (9) left/right.
3. Adjusting the condenser of the Epi-fluorescence device:
B (blue)
G (green)
O(ordinary transmission light)
Coaxial coarse and fine focusing mechanism; Double
layers mechanical stage; ABBE condenser; Built –in
halogen lamp with brightness adjustor. Koehler
illuminating system.
Articulated free binocular head; or Articulated free
trinocular head; or sliding binocular head; or sliding
trinocular head.
Wide field plane-scope eyepiece :
WF10X/18;WF16X/13
Achromatic objective for fluorescence device:
4X;10X;40X(s);100X(s, oil)
In: AC90-254V; Out: 100W DC; Fuse: 5A
HBO Ultra-high pressure mercury
Mercury lamp
spherical lamp: 100W
Halogen lamp
6V/20W Halogen lamp; Fuse: 1A
Special immersion oil for fluorescence device
Specification
EX490; DM510; BA515
EX545; DM580; BA590
P.4
Quantity
1Set
1Set
1Set
1Pair/each
1Set
1Set
2Pcs
2Pcs
1Pcs
1bottle
1Pcs
1Copy