Przeglądaj online lub pobierz pdf Biuletyn serwisowy dla Jednostka sterująca BD Alaris 8000. BD Alaris 8000 14 stron. Inspecting the pump module bezel
2. There are two variations of the bezel that can be differentiated by the lower door hinge:
one variation has a notch on the lower door hinge and the other does not. To identify
which bezel your pump module has:
a. Open the pump module door and hold down the platen to view the lower door hinge.
If the lower door hinge has a notch, the recall does not affect this pump module.
Mark the pump module as unaffected by the recall. If the lower door hinge does not
have a notch, proceed to Inspecting the Bezel for External Damage on page 6.
NOTE:
Alaris pump modules, manufactured prior to April 2011, do not have a notch in the lower door hinge.
If these pump modules are encountered and have not been serviced with customer-ordered bezel
assembly kits P/N 10964559 or P/N 49000204, the recall does not affect these pump modules.
No notch
(recalled)
PN P00000360
Record number 1103-001-008-R-SD-INF, DIR 10000307230, Version 01 Release Date: 16AUG17
DOCO Use Only (Approver's List): 0207-001-001-S (Latest Version) to determine authorized approvers
DOCO Use Only (Record Format Template) 0204-011-005-R, DIR 10000153640, Ver 02, Release Date: 04APR16
Inspecting the Alaris™ Pump Module Bezel
Platen
DIR No.: 10000351665 Version: 00
Notch
(not recalled)
Page 5 of 14