3.7 Installation steps for G 1/8" internal thread (DMD 341)
- Seal the pressure ports of the differential pressure
transmitter in a way that is suitable for your application.
(seals are not included in the scope of delivery)
- Screw the fittings into the threads as far as possible
- Tighten the fittings properly (max. 10 Nm).
3.8 Installation steps for tube nozzle ∅ ∅ ∅ ∅ 6.6x11 (DMD 341)
- Slip your flexible tubes (∅ 6 mm) onto the tube nozzles
as far as possible.
4. Electrical installation
U U U U
WARNING! Install the device only when depressurized
and currentless!
Establish the electrical connection of the device according to
the technical data shown on the manufacturing label, the
following table and the respective wiring diagram.
Pin configuration
Wiring diagrams
2-wire-system (current)
supply +
p
A
I
supply –
3-wire-system (current/supply)
supply +
p
supply –
A / V
I/U
signal +
!
For a device with ISO 4400 plug and cable socket, you
have to make sure that the external diameter of the
used cable is within the allowed clamping range.
Moreover you have to ensure that it lies in the cable
gland firmly and cleftlessly! To ensure the ingress
protection mentioned in the data sheet, the socket has
to be mounted properly. Please check if the delivered
seal is placed between plug and cable socket. After
connecting the cable fasten the cable socket on the
device by using the screw.
For the electrical connection a shielded and twisted
multicore cable is recommended.
5. Initial start-up
Please note that for starting up, the device has to be
stressed by pressure simultaneously at both pressure ports.
Otherwise the sensor could be damaged. For one-sided
pressure admission, the permissible static pressure (one-
sided) must be attended. Please take this out of the current
data sheet.
U U U U
WARNING! Before start-up, the user has to check for
proper installation and for any visible defects.
U U U U
WARNING! The device can be started and operated by
authorized personnel only, who have read and under-
stood the operating manual!
U U U U
WARNING! The device has to be used within the
technical specifications, only (compare the data in the
data sheet)!
6. Placing out of service
U U U U
WARNING! When dismantling the device, it must
always be done in the depressurized and currentless
condition! Check also if the medium has to be drained
off before dismantling!
U U U U
WARNING! Depending on the medium, it may cause
danger for the user. Comply therefore with adequate
precautions for purification.
7. Maintenance
In principle, this device is maintenance-free. If desired, the
housing of the device can be cleaned when switched of
using a damp cloth and non-aggressive cleaning solutions.
Depending on the measuring medium, however, the dia-
phragm may be polluted or coated with deposit. If the
medium is known for such tendencies, the user has to set
appropriate cleaning intervals. If necessary, it is recom-
mended to send the device to BD SENSORS for cleaning.
Please read therefore the chapter "Service/Repair" below.
!
An incorrect cleaning can cause irreparable damages
on diaphragm. Never use spiky objects or pressured air
for cleaning the diaphragm.
8. Service / Repair
8.1 Recalibration
During the life-time of a transmitter, the value of offset and
span may shift. As a consequence, a deviating signal value
in reference to the nominal pressure range starting point or
end point may be transmitted. If one of these two phenom-
ena occurs after prolonged use, a recalibration is recom-
mended to ensure furthermore high accuracy.
8.2 Return
Before every return of your device, whether for recalibration,
decalcification, modifications or repair, it has to be cleaned
carefully and packed shatter-proofed. You have to enclose a
notice of return with detailed defect description when
sending the device. If your device came in contact with
harmful substances, a declaration of decontamination is
additionally required. Appropriate forms can be downloaded
from our homepage www.bdsensors.com. Should you
dispatch a device without a declaration of decontamination
V
S
and if there are any doubts in our service department
regarding the used medium, repair will not be started until an
acceptable declaration is sent.
If the device came in contact with hazardous
substances, certain precautions have to be
complied with for purification!
9. Error handling
V
S
Malfunction
wrong connected
no output
line break
signal
defective ampere meter (signal
input)
load resistance too high
analog output
supply voltage too low
signal too low
defective energy supply
shift of output
diaphragm is contaminated or
signal
damaged
electrical connection is damaged
wrong or no
reverse polarity of the pressure
output signal
ranges
If you detect an error, please try to eliminate it by using this table or send the device to our service address for repair.
!
Improper action and opening can damage the device. Therefore repairs on the device may only be executed by the
manufacturer!
10. Disposal
The device must be disposed according to the
European
Directives
2003/108/EG (on waste electrical and electronic
equipment) Waste of electrical and electronic
equipment may not be disposed by domestic
refuse!
U U U U
WARNING! Depending on the measuring medium,
deposit on the device may cause danger for the user
and the environment. Comply with adequate precau-
tions for purification and dispose of it properly.
11. Warranty conditions
The warranty conditions are subject to the legal warranty
period of 24 months from the date of delivery. In case of
improper use, modifications of or damages to the device, we
do not accept warranty claims. Damaged diaphragms will
also not be accepted. Furthermore, defects due to normal
wear are not subject to warranty services.
Possible cause
inspect the connection
inspect all line connections necessary to supply the device (including
the connector plugs)
inspect the ampere meter (fine-wire fuse) or the analog input of the
PLC
verify the value of the load resistance
verify the output voltage of the power supply
inspect the power supply and the applied supply voltage at the
device
please send the device to BD SENSORS for repair
check the connections
check if the higher pressure is connected with the input "p+"
2002/96/EG
and
Error detection / corrective
12. Declaration of conformity / CE
The
delivered
device
fulfils
all
The applied directives, harmonised standards and docu-
ments are listed in the EC declaration of conformity, which is
available online at: http://www.bdsensors.com.
Additionally, the operational safety is confirmed by the CE
sign on the manufacturing label.
legal
requirements.
DMD_E_011013