Andatech AlcoSense PRODIGY II Podręcznik użytkownika - Strona 4

Przeglądaj online lub pobierz pdf Podręcznik użytkownika dla Sprzęt testowy Andatech AlcoSense PRODIGY II. Andatech AlcoSense PRODIGY II 12 stron. Industrial breathalyser

Andatech AlcoSense PRODIGY II Podręcznik użytkownika

CONDITIONS FOR TESTING

MEASUREMENTS & CONDITIONS FOR TEST
Use a clean mouthpiece for each test.
The optimum temperature range for operating the
unit is -10°C to 50°C. Use within this temperature
range will limit condensation from the subject's
breath and ensure an accurate reading is taken.
Do not allow samples containing smoke, tobacco or
other chemical products to enter the unit, as damage
to the fuel cell can occur.
It is recommended that the operator observes
the subject for approximately 20 minutes before
administering the breath test. This is to ensure the
subject does not consume anything. Following this
procedure will ensure the subject's saliva is clear of
alcohol so that it does not alter the outcome of the
breath sample test.
There must be an interval of at least 15 minutes after
alcohol consumption as alcohol residue in the mouth
may give a false reading. Such residue may also be
left by aromatic drinks, mouthwash, medicines and
drops, as well as by belching and vomiting. Mouth
rinsing with water or non-alcoholic drinks does not
substitute for an interval.
A person's blood alcohol concentration (BAC) can

OPERATING INSTRUCTIONS

ATTACHING A SAMPLING CUP
1. Tear the perforated section from the wrapper of a
new sampling cup.
2. Pull back the wrapper to expose the bottom half of
the sampling cup.
3. Ensuring the two small holes in the bottom of the
sampling cup are aligned with the sensor, insert
the notched end of the sampling cup into the slot
on the right of the screen.
4. Lower the other end of the sampling cup until it
locks into place.
5. Remove remainder of wrapper from sampling cup
and discard.
6. To discard the sampling cup, pull it out in the
direction of the arrow shown on the sampling cup.
5
continue to rise for up to 2 hours after drinking
alcohol. If a result is close to the designated BAC, it
is recommended to wait and retest after 20 minutes.
It can take 10 hours or more for a person's BAC to
return to zero after a high BAC has been reached. A
further test should be carried out later in the day or
the following morning.
LIABILITY FOR PROPER FUNCTION
OR DAMAGE
The liability for proper functioning of the
instrument is irrevocably transferred to the
owner or operator in the event the instrument is
serviced or repaired by personnel not employed
or authorised by Andatech. In such an event, the
warranty will be void.
Working temperature range of Andatech
-10°C to 50°C
Storage temperature range of Andatech
-25°C to 70°C (15~90%RH)
Note: Inappropriate storage of the device will shorten
the lifespan of the fuel cell sensor.
ATTACHING A MOUTHPIECE
1. Tear the perforated section from the wrapper of a
new mouthpiece.
2. Pull back the wrapper to expose half the
mouthpiece.
3. Ensuring the two small holes in the bottom of the
mouthpiece are aligned with the sensor, insert the
notched end of the mouthpiece into the slot on the
right of the screen (refer to Image 1).
4. Lower the other end of the mouthpiece until it
snaps into place (refer to image 2).
5. Remove remainder of wrapper from mouthpiece
and discard.
6. To discard the mouthpiece after a test, slide the
remove mouthpiece button upwards (refer to image
3). This will eject the mouthpiece from the unit.
7. Use a new mouthpiece for each test.
Prodigy II:
®
Prodigy II:
®