Atlantic Ultraviolet Sanidyne Instalacja, obsługa i konserwacja - Strona 8
Przeglądaj online lub pobierz pdf Instalacja, obsługa i konserwacja dla Akcesoria Atlantic Ultraviolet Sanidyne. Atlantic Ultraviolet Sanidyne 16 stron. Portable area sanitizer
Również dla Atlantic Ultraviolet Sanidyne: Podręcznik użytkownika (12 strony)
D
t
etermining
reatment
Using Figure 3 - "Recommended Treatment Time for Room Sizes"
Treatment times are based on a calculated minimum ultraviolet dose of 30,000 microwatt-seconds per square
centimeter at 254nm.
1.
Along the horizontal axis, "Longest Room Dimension (Feet)", locate the room's longest dimension.
2.
Follow up the vertical line until it intersects the heavy black curve.
3.
From the intersection follow back along the horizontal line to the vertical axis, "Timer Setting In Minutes", note the
recommended treatment time.
note: the Sanidyne
is designed to provide direct germicidal ultraviolet exposure for air and surface disinfection in
®
unoccuPied space to disinfect areas and surfaces from microorganisms. Consult our Application Specialists if you
have any questions or need help with the Sanidyne
180
150
120
90
60
30
t
C
reatment
yCle
1.
When the treatment time has been determined, set the unit on a cleared table or on the floor in the center of the room
to be treated. (Treatment is most effective when the unit is in the vertical center of the room and the ultraviolet rays are
unobstructed.)
2.
Post the warning sign(s) outside the entrance(s) to the area being treated and notify the appropriate persons that the area
should not be entered until the disinfection cycle is complete.
3.
Connect power to the unit and the red Power indicator should light.
4.
Remove the protective storage cover.
Set the required time using the hourS and MinuteS up and down buttons. When the Treatment time is displayed
5.
correctly, the Start button can be pushed. An audio alarm will start beeping and the power indicator light will flash as a
warning.
6. Pressing the Start button again will pause the timer and allow a change of the time settings if required. Push again to reset
and restart the exit time.
7. Leave area quickly (within three minutes) to avoid ultraviolet exposure.
8. After the three minute exit period, the beeper will stop and the treatment time will commence.
9. At the start of the last remaining minute of the treatment time (when the treatment time indicates 00:00), the pulsing audible
alarm and flashing light will start again and will stop when the treatment time ends.
warning: Wear personal protection equipment; Gloves, long sleeve shirt with no gaps between
cuffs and gloves and an ultraviolet blocking face shield must be worn to cover and protect eyes and
all exposed skin.
If it is necessary to re-enter the treatment area before the end of the treatment cycle, protective clothing, gloves
and ultraviolet blocking face shield MuSt be worn!
NOTE: Following the end of a treatment cycle, there is a delay of up to two minutes before a fresh treatment
cycle can be started.
t
ime
®
Figure 3 - Recommended Treatment Time for Room Sizes
0
10
20
LONGEST ROOM DIMENSION (FEET)
Based on 8 Foot Ceiling Height
30
40
50
60
8