Do not use cleaning products containing the following ingredients:
AMMONIUM / TRICHLOROETHYLENE
DICHLOROETHYLENE
AMMONIUM CHLORIDE
CHLORINATED or AROMATIC HYDROCARBONS
METHYLENE CHLORIDE
KETONES.
These aggressive agents may damage the plastic parts and cause the device to malfunction.
Please also beware of ALCOHOL-BASED SPRAYS (20% - 40% alcohol content), which may cause the
synthetic materials to tarnish and splinter.
Avoid using iodine solutions capable of irreversibly staining certain clear plastic parts.
The use of disinfectants applied using SPRAYS is ideal for routine disinfection operations and must take
place in compliance with the manufacturer's recommendations and at a distance of 30 cm from the
device, while ensuring that no liquid product is accumulated on the device.
For further information, please contact your company's appropriate department for the provision of
suitable cleaning and disinfection agents.
5.3
Re-installing the components
Once the parts are completely dry, re-insert the two components in the opposite manner to the
disassembly method, presenting each part opposite its tightening clips. The mechanism does not
require a specific forwards/backwards direction of installation. The plate however must be presented
with the fastening hooks facing the rear of the device. Visually check the correct installation of the
different elements.
Ensure that the mechanism is correctly installed facing its clips. Sideways installation could damage
the clips.