Ankom A2000 Instrukcja wymiany
Przeglądaj online lub pobierz pdf Instrukcja wymiany dla Przyrządy pomiarowe Ankom A2000. Ankom A2000 2 stron. Packing and agitator replacement
Również dla Ankom A2000: Procedura serwisowa (2 strony), Procedura serwisowa (3 strony), Procedura serwisowa (3 strony)
1. Unplug the Instrument. Place cardboard
on a table. Lay the A2000 on it's front.
Remove the nuts that secure the back
panel. Remove the panel. Remove the
feet and mounting screws and then
remove the bottom panel.
3. Remove the three nuts that secure
the motor bracket in place.
ANKOM Technology 2052 O'Neil Rd Macedon, NY 14502 PH: (315) 986-8090 FAX: (315) 986-8091
Motor Assembly Replacement
A2000/A2000I
2. Loosen the bottom set screw on
the coupler.
4. Unplug the motor wires.
replacing the assembly, install the new
motor assembly, tighten the mounting
nuts and coupler set screw.
120v Wiring-black to brown,
white to blue
220v Wiring– blue to blue,
brown to brown
5. Replace the back and bottom panel and
the instrument is ready for use.
023
Service Procedure
(Page 1 of 2) 08/06/2014
If you are