Disassemble the valve.
2.
a. Turn the instrument power switch to the OFF position and
disconnect the power cord from the power source.
b. Remove the two screws from the bottom of the port A
valve.
c. Slide the Coil off of the valve.
d. Remove the valve stem from the valve being careful not to
lose the spring in the stem.
Clean the Teflon seal on the bottom of the valve stem and the
3.
interior of the valve body with hot soapy water.
Re-assemble the valve.
4.
a. Slide the valve stem back into the valve body (be sure the
spring is in the valve stem).
b. Slide the Coil back over the valve stem.
c. Insert the two screws back in the bottom of the valve and
tighten.
Repeat steps 2-4 for the port B valve.
5.
Reconnect the power cord to the power source.
6.
ANKOM Technology  2052 O'Neil Road  Email: [email protected]  Phone: 315-986-8090  Fax: 315-986-8091
A and B Valve Cleaning
A2000/A2000I
Service Procedure
Revised: 08/05/2014
Coil
Screws
004
Valve
Valve
Stem
Teflon
Seal