13. The lid should remain locked until the rotor has nearly stopped. If
the machine unlocks prematurely, contact Drucker Diagnostics for
assistance.
14. Once the rotor has stopped, the interlock system will become
disengaged for sixty (60) seconds. The 'UNLOCKED' indicator light
will illuminate during this time.
15. To gain entry into the centrifuge after this period has ended,
simply press the 'OPEN / STOP' button. The lid will unlock for
fifteen (15) additional seconds.
After the centrifuge has passed this procedure, it is ready for operation.
C
P
ONTROL
ANEL
'RUNNING'
Lights up when the
machine
is in operation, (power is
being applied to the
motor).
'LATCHED'
Lights up when the lid has
been closed and latched
properly.
'UNLOCKED'
Lights up to indicate that
the locking mechanism
has been deactivated,
allowing access to the
rotor chamber.
Page | 7 Drucker Diagnostics +1-814-692-7761 - [email protected]
'START'
Begins a new run, (the lid
must be closed, see pg. 6).
'OPEN/STOP'
Allows for access into the
rotor chamber by
disengaging the locking
mechanism. Entry is only
permitted when the rotor is
stopped. Pressing this
button during operation will
terminate the run and
unlock the lid after the rotor
has come to a stop.