The rotor and accessories are rated for the maximum rotation frequency shown in the table below.
Tube Capacity
Dimensions (H x W x D)
Weight
Sound Level
Environmental Range
Voltage
Frequency
Power Requirement
Centrifuge Motor
Max g-Force
Max Speed
Cycle Time
C
ALCULATING THE
The I.F.U.s of tube manufacturers recommend cycles at a minimum G-Force, which can be calculated if you know
the RPM and the radius. Use the formula below or go to www.druckerdiagnostics.com/g-force-calculator/.
In Centimeters:
RCF or G-force = 0.00001118 x
Rotor Radius (cm) x (RPM)
Radius
Drucker Diagnostics – Customer Service: +1-814-692-7661 – [email protected] | Page 13
HORIZON 12 Flex
12 tubes – 3 to 10 mL
13 in x 15 in x 9 in
(33 cm x 38 cm x 23 cm)
34 lbs. (15 kg)
56 dB A
16 – 32 degree C
95 -253 VAC
50/60 Hz
280 Watts
½ H.P. Brushless
2,000 xg
3,700 RPM
.5 to 99 minutes (+/- 2%)
G-F
ORCE
2
HORIZON 12 Flex
5.25 in (13.3 cm)
HORIZON 24 Flex
24 tubes – 3 to 10 mL,
12 tubes – 15 mL
6 tubes - 50 mL
15 in x 17 in x 9 in
(38 cm x 43 cm x 23 cm)
39 lbs. (17 kg)
59 dB A
16 – 32 degree C
95 -253 VAC
50/60 Hz
280 Watts
½ H.P. Brushless
2,000 xg
3,400 RPM
.5 to 99 minutes (+/- 2%)
In Inches:
RCF or G-force = 0.0000284 x
Rotor Radius (in) x (RPM)
HORIZON 24 Flex
6 in (15.3 cm)
2