Drucker Diagnostics 642B Instrukcja obsługi - Strona 6
Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Sprzęt diagnostyczny Drucker Diagnostics 642B. Drucker Diagnostics 642B 16 stron.
S
L
ETUP
OCATION
1. Unpack the centrifuge and verify that all of the supplied
equipment is present.
2. Choose a setup location which meets the following criteria:
a.
A clearance height of 20" is required to open the lid.
b. The clearance envelope is the space around the centrifuge
which is required for safety. Choose a setup location which
will allow for a clearance envelope of at least 24" x 24",
(with the centrifuge at the center). No person or hazardous
material shall be permitted in the clearance envelope during
operation. The operator time within the envelope shall be
limited to the time necessary for loading, unloading and
centrifuge operation only.
c.
Proper ventilation is necessary to prevent the overheating of
samples as well as premature failure of the centrifuge.
Choose an area which will allow unencumbered air flow.
d. The centrifuge is designed to secure to the operating surface
by four suction feet. No adjustment is necessary for leveling
the centrifuge; however, the surface should be flat and level.
e. Be sure the outlet is always within reach as the line cord is
the means of emergency disconnection!
Drucker Diagnostics +1-814-692-7761 - [email protected] Page | 4