Drucker Diagnostics Horizon 6 FA Instrukcja serwisowa - Strona 5

Przeglądaj online lub pobierz pdf Instrukcja serwisowa dla Sprzęt diagnostyczny Drucker Diagnostics Horizon 6 FA. Drucker Diagnostics Horizon 6 FA 10 stron.

Drucker Diagnostics Horizon 6 FA Instrukcja serwisowa
PROBLEM
Cycle ends prematurely:
Centrifuge reports the error in a repeating BUZZ / BEEP sequence
ROTOR BALANCE ERROR:
4 error buzzes + 2 repeating beeps
All LEDs blinking in unison with beeps
OVERCURRENT ERROR:
4 error buzzes + 3 repeating beeps
MOTOR SPEED ERROR:
4 error buzzes + 4 repeating beeps

7. SERVICE INSTRUCTIONS

7.1. Cleaning
The cabinet, rotor and accessories can be cleaned using soap and water, isopropyl
alcohol, or a 10% (5500 PPM) bleach solution
Apply cleaning solutions with a towel or cloth. Do not submerge the centrifuge in
water or other cleaning solutions as this will cause damage and void the warranty.
Under no circumstances should any of the following be used: TBQ Germicidal
Products, Fully/Partially Halogenated Hydrocarbons, Ketones or Esters.
Use of any chemicals not prescribed by the manufacturer may cause damage to the
rotor and tube carriers and shall not be used.
7.2. Removing the Rotor
Use a 1/2" nut driver to loosen the center rotor screw (turn counter-clockwise).
Lift the rotor straight up and out of the rotor chamber.
To install the rotor, reverse steps A and B above.
Take care to align the hub spines with the rotor hub.
Tighten the rotor nut by hand with a ½" nut driver until snug. A good rule of thumb
is one full thread showing above the nut is a good indicator of proper installation.
7.3. Maintaining the Rotor
Keep the rotor clean, any corrosive materials must not be allowed contact with the
rotor and should be cleaned immediately.
The rotor should be checked periodically for signs of wear.
Remove the rotor from service if any of the following are found: cracks, deep
scratches, corrosion or discoloring.
DRUCKER DIAGNOSTICS
HORIZON 6 SERVICE MANUAL
POSSIBLE CAUSE
Rotor improperly loaded
Control board, motor, or wiring
fault
Rotor motion not detected
Abnormal rotor acceleration
detected
Motor fails to reach the set
speed
SM024
REV A
SOLUTION
Correct the imbalance
Check wiring
• Check sensor in lid locking tray
assembly
• Check reflective tape on rotor
Check for missing tube holders
• Check for missing tube holders
• Check the lid and guard bowl
gaskets. Voids in the rotor
chamber gaskets change the
airflow, resulting in increased
stress on the motor drive
5