DS Europe AN-601 Series Instrukcja obsługi - Strona 7

Przeglądaj online lub pobierz pdf Instrukcja obsługi dla Przyrządy pomiarowe DS Europe AN-601 Series. DS Europe AN-601 Series 11 stron. Portable instrument

1 AN601 batteries

The AN601 instrument is to be powered with 4pcs AA 1,5V batteries. The operational autonomy of the
instrument will depend on the battery type. It is advisable to use rechargeable batteries for digital
cameras with high charge capability like 2600mA/hour or higher.

1.1 Instrument autonomy of operation

The AN601, when connected to a 350 Ohm strain gauge instrument, is consuming approximately
60mA.
The battery autonomy is directly proportional to their charge status and if it is used a 2600mA/Hour
type then the autonomy is: 2600mA/hour/60mA=40 hours of continuous working.

1.2 Battery housing

Detail A
Push this part to open the lid, for opening the battery housing.
Fig. 1 : AN601 battery housing.

1.3 Battery status led

A red led switches on when the batteries have low charge status.
It is advisable to change the batteries when they are no more charged in order to avoid wrong readings
on display (if batteries are rechargeable then recharge them).
Note:
The AN601 recognizes a battery low charge status by using memorized parameters referring to NiMa
types with 2600mA, different models could change the activation of the signaling led.
MANUALE PRELIMINARE AN 601 V1 REV. 0 dated 09/10/07
-7-
To open the battery housing, on AN601 rear side, it
is necessary to push upwardly the plastic lid and
then lift it.
Insert batteries with correct polarity.
If batteries are inserted with wrong polarity,
instrument is not damaged.
After having inserted batteries then close the lid.