Immunity Test 
Conducted RF 
3 Veff from 150KHz to 
EN 61000‐4‐6 
 
80MHz 
 
 
 
 
Radiated RF 
3 Veff from 80MHz to 
IEC 61000‐4‐3 
 
2.5GHz 
 
 
 
 
 
 
 
Field strengths from fixed RF transmitter, as determined by an electromagnetic site survey 
level in each frequency range
 
 
Recommended separation distance between portable and mobile RF communications equipment and the B‐ONE 
The B‐ONE is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or 
the user of the B‐ONE can help prevent electromagnetic interference by maintaining a minimum distance between portable and 
mobile RF communications equipment (transmitters) and the B‐ONE as recommended below, according to the maximum output 
power of the communications equipment.  
Rated maximum 
output power of 
transmitter (W) 
 
From 150kHz to 80MHz 
0,01 
0,1 
10 
100 
 
For transmitters rated at a maximum output power not listed above, the recommended separation distance (D) in meters (m) can be 
estimated  using  the  equation  applicable  to  the  frequency  of  the  transmitter,  where  (P)  is  the maximum  output  power  rating  of  the 
transmitter in watts (W) according to the transmitter manufacturer. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
USER MANUAL Rev 25/05/2023 
 
Test level 
Compliance level  
IEC 60601‐1‐2 
3 Veff from 150KHz 
to 80MHz 
 
 
3 Veff from 80MHz 
to 2.5GHz 
 
 b
. Interference may occur in the proximity of equipment marked with the following symbol:  
Separation distance according in frequency of transmitter (m) 
From 80MHz to 
d= 1,2P 
800MHz 
0,12 
0,38 
1,2 
3,8 
12 
ATS DENTAL srl – 51034 Serravalle Pistoiese – Italy – www.atsdental.it – [email protected]  
B‐ONE 
Page 14 of 16 
Electromagnetic Environment ‐ Guidance 
Portable and mobile RF communication equipment 
should be used no closer to any part of the B‐ONE, 
including cables, than the recommended separation 
distance calculated from the equation applicable to the 
frequency of the transmitter. 
Recommended separation distance 
d = 1,2P from 150KHz to 80MHz 
d = 1,2P from 80MHz to 800MHz 
d = 2,3P from 800MHz to 2.5GHz 
where P is the maximum output power rating of the 
transmitter in watts (w) according to the transmitter 
manufacturer and recommended separation distance in 
metres (m). 
 
From 800MHz to 2,5GHz 
d= 1.2P 
0,12 
0,38 
1,2 
3,8 
12 
 
 
 
a
 should be less than the compliance 
d= 2,3P 
0,23 
0,73 
2,3 
7,3 
23