ATS Dental B-ONE Instrukcja obsługi i konserwacji - Strona 8

Przeglądaj online lub pobierz pdf Instrukcja obsługi i konserwacji dla Sprzęt medyczny ATS Dental B-ONE. ATS Dental B-ONE 17 stron.

 
 
 
 
"REF":  BNE300 
 
 
 
 
 
"DELUXE" 
 
 
 
 
1.
Open the pump Insert the rotor of the pump inside the pump. Close the pump. 
2.
Hang the bottle or bag of physiological saline from the bracket. 
3.
Insert the irrigation line perforating into the bottle or bag of physiological saline. 
4.
Connect the end of irrigation line to the contrangle. 
 
NB: Use only ATS DENTAL Tube. Do not run the pump without tube. 
Do not use coolant reservoirs that can hold more than 1 liter. Ensure that the device is standing securely. Use transparent coolant 
reservoirs. 
 
 
 
 
 
 
 
 
 
 
 
 
 
USER MANUAL Rev 25/05/2023 
 
7. PEDAL (FOOT CONTROL OPERATION) 
 
 
 
 
8. IRRIGATION LINE SETUP 
ATS DENTAL srl – 51034 Serravalle Pistoiese – Italy – www.atsdental.it – [email protected]  
B‐ONE 
Page 8 of 16 
The  left‐hand  button  allows  user  to  change  the  value  of  the 
coolant pump as well as the key "flow" on the keyboard. 
The central up button allows user to change the programs as 
well as the key "P ‐ programs" on the keyboard. 
The right‐hand button allows user to switch between forward 
and reverse motor rotation (when in reverse, console will emit 
a beeping tone), as well as the key "Fwd/Rev" on the keyboard. 
The central down button allows user to control up to the pre‐
selected maximum speed and switch on the pump if enabled. 
 
 
 
 
 
 
 
 
 
 
NB: Only in case of using a new injector, 
and only if a reduced flow of the coolant 
is felt, insert 2 drops of SILICONE OIL 
(easily available on the market) in the 
center of the rotor to limit friction.