User Maintenance

Warning: Prior to any maintenance or cleaning of this unit, disconnect the power cord from
the power supply.
Avertissement: Avant d'effectuer toute maintenance ou entretien de cet appareil, débrancher
le cordon secteur de la source d'alimentation.

Cleaning and Disinfecting

If a hazardous material or substance have spilled, immediately initiate your site's Hazardous Material
Spill Containment protocol. Contact your local Site Safety Officer and follow instructions per the site
policy and procedures.
Do not use cleaners or disinfectants that contain solvents capable of harming paint coatings
or stainless steel surfaces. Do not use chlorine-based bleaches or abrasives; these will damage
the COLDWELL liner.

Vacuum Pump Maintenance

Vacuum pumps are high wear equipment and require replacement of seals and diaphragms at least
once per year. Poor vacuum performance or bubbling or strained noises from the pump may indicate
that the pump is need of maintenance.
Chemical duty diaphragm pumps commonly require diaphragm and seal replacement
every 6 to 8 months.
Dry scroll pumps typically require replacement of the tip seals every 9000 hours of use or
at least once per year.
Pump Manuals
Refer to the operation manual supplied with your vacuum pump for recommended maintenance rou-
tines such as oil levels, replacement of sorbent charge, and exhaust filter change outs. Contact your
vacuum pump supplier if you do not have an operation manual.