3Shape L1P-1F Instrukcje bezpieczeństwa i konfiguracji - Strona 4
Przeglądaj online lub pobierz pdf Instrukcje bezpieczeństwa i konfiguracji dla Sprzęt dentystyczny 3Shape L1P-1F. 3Shape L1P-1F 12 stron.
4.7 TRIOS Scanner, TRIOS Scanner Tip and TRIOS
Protection Tip Description
TRIOS system consists of:
Please follow the steps below for attaching and detaching the scanner tip and protection tip:
The TRIOS scanner has built-in tip detection.
A set of TRIOS scanner tips are delivered with the TRIOS scanner.
The TRIOS scanner tip is the only applied part.
IMPORTANT NOTICE
The TRIOS scanner tips can only be mounted one way
4.8 Battery and Battery Charger Description
Power (mains) cable and plug:
The charger has a detachable cable (C8 connector) appropriate for your location, supplied
in the packaging.
4.9 Other System Elements Description
For L1P-1F systems:
TRIOS Patient Monitoring (TPM) Software:
Software for aid in diagnostics of caries.
4
TRIOS Scanner (L1-2):
Wireless handheld dental intraoral scanner for
scanning a patient.
Two main areas of the TRIOS scanner are referred
to in this guide: mirror housing and scanner body.
TRIOS Scanner tip (TST-11):
Re-usable TRIOS scanner tip attached to the TRIOS
scanner. A tip is needed for the scanner to be able
to scan.
TRIOS Sleeve (TSL-01):
Single use sleeve to cover the TRIOS scanner.
TRIOS Protection Tip (TST-13):
Protects the optical window of the TRIOS scanner
from damage and dirt when the TRIOS scanner
is not in use. Enables the scanner to run self-test,
when mounted.
1. To attach the tip, place the tip over the mirror
housing facing downwards until it clicks.
2. To detach the tip, pull the tip away from
the mirror housing.
A set of batteries for the TRIOS scanner (TB-04):
Batteries for the TRIOS scanner with LED light
to indicate battery status.
Be aware that the batteries must be stored and
operated under correct conditions, see section 4.13.
A battery charger (C2BP):
With two ports to charge TB-04 batteries.
Be aware that the battery charger must be stored
and operated under correct conditions,
see section 4.13.
Pod (LP1-2):
Holder for the scanner.
A USB wireless dongle for the PC.
A USB license dongle.
4.10 Setting up TRIOS
It is recommended that unpacking, assembly and connection of the TRIOS System be performed
by TRIOS authorized technical service personnel.
It is recommended to fully charge batteries for the TRIOS scanner before use.
CAUTION
Let the system adjust to room temperature before use.
WARNING
Examine all parts of the system. Look for physical damage, loose parts
or signs of wear that could interfere with proper use and functionality.
Contact your service provider if there is visible damage or malfunctioning.
Charging batteries:
Before you start:
Place the battery charger on a flat surface horizontally.
Step 1: Attach the power cord to the charger and plug it into mains power.
Step 2: Push the batteries one by one into the charger until they click into place.
When charging, the 5 LEDs on the end of the battery will indicate the status:
Each LED represents 20% of the battery capacity.
Five constant LEDs means that the charging
is complete.
While charging is ongoing each LED will light up one
at a time. The battery status will then be shown, with
each LED representing 20% battery capacity.
If any other charging pattern appears, a charging
error has occurred. In this case, remove the
batteries from the charger and reinsert them.
If charging errors continue, please contact 3Shape
support ([email protected]).
There is an LED under the battery charger. The LED has the following indications:
A green light indicates the battery charger is on
and working
A blue light indicates the power is unstable.
Unplug and reinsert the power for the charger.
If the LED continues to indicate a blue light
please contact support ([email protected]).
A white light indicates the battery may have been
inserted incorrectly. Try to insert the battery again.
If the LED continues to indicate a white light,
please contact support ([email protected]).
Please follow the steps below for assembling the TRIOS system:
Before you start: Place the pod on a flat and horizontal surface optimally in relation
to the patient chair and your work position. Attach the TRIOS protection tip to the
TRIOS scanner if not already in place.
Step 1: Insert a fully charged battery into the rear
of the TRIOS scanner.
Step 2: Place the TRIOS scanner in the pod.
Step 3: Clean and sterilize a TRIOS scanner tip prior to first time usage.
See section 7 for cleaning and sterilization instructions.